10W adapter for medical equipments,Website:http://www.szhehuiyuan.com, Output 5V/2A, DoE level VI; different plugs options for different countries's standard.Features:Electrical Specifications:Model No.CC00332EfficiencyDoE Level VI and CoC V tier 2Input Voltage100~240Vac (90~264Vac min/max)No Load Input Power≦75mW @ 230VAC inputFrequency50/60Hz Nom (47~63Hz min/max)ProtectionOVP, OCP, SCP, OTPInput Current0.3A max at 100VacInsulation Resistance>100M@ 500VdcOutput Voltage5VDielectric WithstandPrimary to secondary, >4000Vac, 1 minuteOutput Current2AOVPOutput voltage shall not exceed 9VMaximum Output Power10WOCP, SCPAuto recoveryOutput Ripple & Noise100mVp-pOTPInclude an over-temperature protectionLeakage Current )<35uA @ 264Vac/50Hz (meet Type BF req.)Average Efficiency79% min.AC Inrush Current60A max. For cold start and at rated inputCable Length1500 mmOver-shoot< 10% of nominal voltage valueDimension (mm)50 x 71.7 x 20Hold-up Time16mS min @ 115Vac/60Hz and 2.0A loadUS, EU and UK,Turn-On Delay Time<3 sec at 2.0A loadEnvironmental Specifications:Safety and Compliance:Operating Temperature5℃ ~ +40℃Safety IEC/UL 60601-1:2005+A1:2012 2xMOPPStorage Temperature-25℃ ~ +70℃IEC60601-1-2 Ed4 2014Humidity, Operation15% ~ 93% RHImmunityHumidity, Storage10% ~ 93% RHESD EN61000-4-2 Criterion AMTBF(MIL-STD-217F) 50K hr min. at 25°CRS EN61000-4-3 Criterion AEFT EN61000-4-4 Criterion BSurge Test EN61000-4-5 Criterion BCS Test EN61000-4-6 Criterion APower Frequency MagneticFieldEN61000-4-8 Criterion AVoltage Dips EN61000-4-11 Criterion BEMI Requirement FCC Class B, EN55022 Class B