Constant temperature and humidity test machine is also called environment testing machine, used to test various types of materials for their tolerances of heat, cold, dry, humidity. Application: Quality inspection for all types of electronic products, electric appliance, and products from communication, instrument, automobile, plastic, metal, food, chemical, building materials, medical and aerospace industries. |
Major Specification
Model |
CZ-E-1000A |
Test specimen limitation |
1)inflammable material ,explosive material, easy to volatile material test sample are forbidden; 2)storage corrosive substances test sample is forbidden; 3)storage biological test is forbidden; 4)storage strong electromagnetic emission source for test sample or storage are forbidden; |
Capacity / Weight / Dimension |
|
Internal capacity |
1000 L |
Internal dimension |
1000×1000×1000(mm) W×H×D |
External dimension |
|
Weight |
|
Working noise |
≤70db (The value got at the point with 1 meter away from the front site of the machine and 1.2 height from the ground.) |
Watt / Power / Current |
|
Maximum watt |
7.5KW |
Maximum current |
9A |
Power supply requirement and source |
1)AC 380V three phases with four wires +protective grounding; voltage fluctuation range 10% 2)Frequency fluctuation range:50±0.5Hz; 3)Power supply method: TN-S or TT method 4)Grounding resistance of protection ground wire<4Ω 5)User should equip with relevant capacity air switch or dynamic switch on the installation site, and the switch must control this machine only. When the electrified test sample is put in the test chamber, the power of the test sample should be connected with extra power, don’t use the power of this machine; |
Performance Index |
|
Environment requirement |
The data got under the condition of environment temperature is +25℃,and R.H. is ≤85%,no test sample in the chamber. |
Temperature range |
temperature range:+25℃~+150℃ |
Control accuracy |
temperature resolution:0.01℃ temperature deviation: ±1℃ temperature uniformity:≤2℃ temperature fluctuation:±0.5℃ |
Heating /cooling rate |
+25℃→+150℃≤45 min +150℃→+25℃≤10 min |
Note: Steel heat capacity ≤35kg/m3 ;when humidity time, no heat resource; |
|
Note: performance index above is under the condition of environment temperature is +25℃, and R.H. is ≤85%,no test sample in the chamber; |
|
Standard and test condition |
1. GB2423.1-2008/IEC6008-2-1-2007 Electronic Electrician Product Environmental Test Part 2:test methods Test A: low temperature 2. GB/T 2423.2-2008 Electronic Electrician Product Environmental Test Part 2:test methods Test B: high temperature 3. GJB150.3A-2009 High temperature test method 4. GJB150.4A-2009 Low temperature test method 5. GB2423.22-2008 Electronic Electrician Product Environmental Test Part 2:test methods 6. GBT 2424.5-2006 Electronic Electrician Product Environmental Test Conformation of performance for temperature test chamber 7.GB/T 10589-2008 Low temperature test chamber technical condition |